S-000021Wrigley
256-channel adult net, 54-56 cm · part 4733030
Needs repairWarranty out
Net status
Turn the net to find a sensor. Drag to rotate, pinch or scroll to zoom.
Over threshold at the last bucket testNo signal at the last bucket testReported and not yet fixed
Open problems (5)
- Missing pedestalsReported
- Electrode 54
Anonymous · 3 weeks ago · October 25, 2026 at 1:52 pm
- Broken electrodesReported
- Electrode 132 — missing entirely
Anonymous · 3 weeks ago · October 25, 2026 at 1:52 pm
Photos - Broken electrodesReported
- Electrode 132 — missing entirely
Anonymous · 3 weeks ago · October 25, 2026 at 9:49 am
Photos - Missing pedestalsReported
- Electrode 54
Anonymous · 3 weeks ago · October 24, 2026 at 6:49 pm
- Broken electrodesWaiting on a bucket test
- Electrode 219 — high impedance219 at 521.8 kΩ
Mani Vahdatizadeh · 5 weeks ago · October 10, 2026 at 7:20 pm
Repaired. A bucket test has to confirm the electrodes before this closes.
Photos
Repaired electrodes
Label
Scanning this opens the report form with this net already chosen.
S-000021
Wrigley
Scan to report a problem
Bucket tests
- October 17, 2026No signal: 254Over threshold: 129, 241, 249Mani Vahdatizadeh · threshold 50 kΩ
- October 17, 2026Every electrode passedMani Vahdatizadeh · threshold 50 kΩ
- September 15, 2026No signal: 246, 254Over threshold: 241, 249Mani Vahdatizadeh · threshold 50 kΩ
- January 1, 2026Every electrode passedThe spreadsheet said “none”Imported · threshold 50 kΩ · from the old spreadsheet
- January 1, 2025No signal: 1, 244, 251Imported · threshold 50 kΩ · from the old spreadsheet
- September 1, 2022No signal: 251, 254Over threshold: 244Imported · threshold 50 kΩ · from the old spreadsheet
Historical notes
- Chin straps need fixing (Jan 2025). Verify 244 works. Potentially damaged during repair
- Disconnected electrodes 31 and 32
- Missing pedestal on 31
- Broken pedestal on 242
- Also, roughly 30 electrodes could not get an impedance below 50
Reported electrodes: 54, 132, 219